for more product details, please call us:0731-84930888
for more product details, please call us:0731-84930888
多功能的直流标准源仪器,可输出高准确度的直流电压、电流、功率及电能
准确度为 0.01 级 / 0.02 级 / 0.05 级可选
电压输出:最大达 1150 v
电流输出:最大达 33 a ( td1510 为 110 a ),小电流低至 0.1 μa
overview
model td1500, a wide-range, high-precision and multi-function calibrator with a built-in dc standard source. the apparatus can accurately outputs dc voltage, current, power and electrical energy with the optional accuracy classes are 0.01, 0.02 and 0.05. applied to calibrate various types of dc electric measuring instruments.
features
● dc electrical parameters output: voltage is up to 1150 v, current is up to 33 a, the low-current is as low as 0.1 μa.
● value output mode: standard source and output source modes, easy to calibrate both digital meters and analogue meters.
● value adjustment method: fixed-point, rotary encoder, step and potentiometer (option).
● output switch: the output can be turned on/off arbitrarily by one-touch operation.
● statistics and analysis of data: calculate the maximum, minimum, average and standard deviation, and then analyze the stability and trends.
● excellent human-computer interaction: large lcd touch screen display, touch and button operation modes, easy for users to measure manually.
● electric energy meter calibration: a built-in dc low-voltage standard source with an output range of ±(10 μv to 4.4 v), and standard pulse i/o interfaces to calibrate electrical energy meters by direct access or indirect access.
● transmitter measurement (option): measure the secondary dc signal of the transmitters.
● proprietary software (option): support the calibration of the checked meters by semi-automatic or full-automatic, data management and report export.
applications (class 0.05)
● calibrate dc voltmeters and ammeters (class 0.2 and below)
● calibrate dc power meters (class 0.5 and below)
● calibrate dc electric energy meters (class 0.5 and below) (option)
● calibrate dc transmitters (class 0.2 and below) (option)
electrical specifications
dc voltage
output characteristics
● output range of voltage: 20 mv to 1100 v (1150 v @ electric energy option)
● resolution: 0.001 % of range
● display: 7 digits
● ripple coefficient: <0.5 %
● protection: short-circuit protection, overload protection.
dc current
output characteristics
● output range of current: 0.1 μa (or 1 μa) to 33 a
● resolution: 0.001 % of range
● display: 7 digits
● ripple coefficient: <0.5 %
● protection: open-circuit protection, overload protection.
dc electric energy (option) and dc power
● output range: combination of dc voltage range and dc current (or low-voltage) range.
● accuracy: addition of dc voltage accuracy and dc current (or low-voltage) accuracy.
● input of standard electric energy pulse:
receive high frequency electric energy pulses (≤200khz), input levels ≤ 24v
● output of standard electric energy pulse:
high frequency full scale @ 60 khz, low frequency full scale @ 6 hz.
frequency jitter: <10 μs
stability of clock frequency in 1 year: <50 ppm
● electric energy measurement error: automatically display with 0.0001 %.
dc small signal measurement (option)
● range: 1 v, 10 v, 2 ma and 20 ma
● measuring range: ± (0 to 12 v);± (0 to 24 ma)
● accuracy: ± (0.006 % of reading + 0.004 % of range)
● notes: applied to measure the secondary dc signal of the transmitters.
general specifications
● line power: 220 v ±22 v, 50 hz ±2 hz
● maximum power consumption: 700 va
● operating environment: 0 °c to +40 °c, 20 % to 85 % r·h, non-condensing
● storage environment: -20 °c to +70 °c, <95 % r·h, non-condensing
● dimensions (w × h × d): 525 mm × 210 mm × 425 mm
● weight: 17.5 kg
● communication: rs-232 interface
selection guide