td4550 portable calibrating apparatus for three-凯发k8国际

for more product details, please call us:0731-84930888

homeelectrical measurementac energy calibrators

td4550 portable calibrating apparatus for three-phase meters

专用于检定三相交流电能表、功率表的便携式仪器


准确度为 0.02 级 / 0.05 级 / 0.1 级可选

三相标准源输出最大 456 v / 120 a,各种负载下均能保证量值输出稳定

电流输出最小达 2 ma,是被检电能表进行起动试验的基础


overview

td4550 is a portable apparatus specially applied to the verification of three-phase meters and wattmeters, the accuracy is 0.02/0.05/0.1 optional. it can accomplish the test items including the basic error test, start-up test, latent test, meter constant, standard deviation, frequency and harmonic effect test and so on.reference standards are gb/t 11150-2001、jjg 597-2005、jjg 307-2006、jjg 596-2012.

features

● the maximum ac output capacity is 456 v/120 a, which can ensure the stability of output value at full load.

● the minimum ac current output to 2 ma, which is the basis for start-up test of the verified meter.

● the magnitude adjusting has methods of rotary encoder, bit adjustment and step ratio. it can be adjusted by three-phase system or phase separation. the three phase parameters can be adjusted uniformly or separately.

● phasor diagram display: the phase between the three-phase electric parameters can be arbitrarily adjusted and displayed through the form of phasor diagram.

● man-machine function: the workbench is equipped with large size and high-definition lcd, which is intuitive and convenient for operation.

 harmonic function (options): it can load 2nd-21st harmonic with adjustable amplitude and phase to complete the harmonic influence test.

● proprietary software: semi-automatic and full-automatic verification is supported, so as data management and certificate export.

applications (class 0.05)

1

2

technical specifications

ac voltage/current output

3

4

output characteristics:

 range of voltage: 0 v~264v

 range of current: 0.1 ma~120 a

● distortion factor: < 0.5%

● resolution: 0.002%*rg

 display: 6 digits

● symmetry of three-phase parameter: voltage≤+ 0.2%; current≤+ 0.5%; phase≤+ 0.5°

● protection: short-circuit protection, open-circuit protection, overload protection

frequency, phase and harmonic

● range of frequency45.000 hz~65.000 hz

   resolution: 0.001 hz

   accuracy: ± 0.01 hz

● range of phase0.000°~359.999°

   resolution0.005°

   accuracy± 0.01°

● harmonic output(option): 2nd to 21st,amplitude 0~40%、phase 0.0°~359.9° adjustable

stability of output power

5

accuracy of power/electrical energy

6

measurement characteristics:

● ac power rangecombination of ac voltage range and current range

● output of standard electrical energy pulse (high frequency): full scale value corresponds to 60 khz.

   output of standard electrical energy pulse(low frequency): full scale value corresponds to 6 hz.

   frequency jitter of output pulse: < 10 μs

● input of standard electrical energy pulse: high frequency pulse less than or equal to 150khz with the amplitude of 3.3v~24v can be received.

● the electrical energy error is automatically displayed with the resolution of 0.0001%.

general specifications

● line power: ac ( 220 ± 22 ) v,( 50 ± 2 ) hz

● operating environment: 0 ℃~40 ℃, 20%~80% r·h,  non-condensing 

● storage environment: -20 ℃~70 ℃,< 80% r·h,不结露

● dimensions(width*depth*height)490 mm × 485 mm × 200 mm

● weight: 29 kg