for more product details, please call us:0731-84930888
for more product details, please call us:0731-84930888
专用于检定直流高压高值电阻器、高压电阻箱、直流高压表
阻值范围:10 ω ~ 10 tω
电压范围: 0 ~ 10 kv
overview
model td1220, an intelligent calibrator for high voltage and value dc resistors, high voltage resistance boxes and high voltage dc meters. the resistance range is from 10 ω to 10 tω, and the voltage output range is from 0 to 10 kv. applied to calibrate the high resistance box with class 0.05 and below.
meet the requirements of jjg 1072-2011 verification regulation of high voltage and value d.c. resistors and dl/t 979-2005 verification regulation of high voltage d.c. resistance box.
applications
● calibrate high voltage and value dc resistors and resistance boxes, high voltage dc meters, milliammeters and voltage dividers.
● calibrate the insulation resistance tester calibration devices
● calibrate the megger calibrators
features
● constant current (cc) mode: the stability of current output is 0.01 %/h, applied to measure low resistance. the maximum voltage on this mode is 1050 v, can prevent high voltage from damaging the low resistance dut.
● constant voltage (cv) mode: the stability of current output is 0.005 %/h, applied to measure high resistance. the current output of this mode is limited to 3 ma or less, can protect the measured resistance and the calibrator itself.
● simultaneously output dc voltage or current and measure resistance, the accuracy of calibrating high voltage dc meters is up to 3×10-4.
● the measurement uncertainty from 100 ω to 100 mω is better than 2×10-4, the measurement accuracy of 10 tω is still up to ±2 %.
● all-digital button output mode, which is better than the mode of rotary shirt switch, makes the error calculation much simpler.
● good electrical isolation and protection ensure high safety and reliability.
● large lcd touch screen, more intuitive of value display and easier operation.
● proprietary software (option): support the calibration of the checked meters by semi-automatic or full-automatic, data management and report export.
electrical specifications
dc voltage/current output
output characteristics
● resolution: 0.002 % of range
● display: 6 digits
resistance measurement
general specifications
● line power: 220 v ±22 v, 50 hz ±2 hz
● maximum power consumption: 80 va
● operating environment: +18 °c to +28 °c, 40 % to 80 % r·h, non-condensing
● storage environment: -20 °c to +70 °c, <80 % r·h, non-condensing
● dimensions (w × h × d): 440 mm × 170 mm × 430 mm
● weight: 11 kg
● communication: rs-232 interface
selection guide