for more product details, please call us:0731-84930888
for more product details, please call us:0731-84930888
适用于检定无源泄漏电流仪 ( 不含隔离供电电源 ) 或有源泄漏电流仪
可完成检测项目:
泄漏电流误差检定
试验电压误差检定
输入电阻检测
输入电路时间常数测量
overview
model td1200, a calibrator for leakage current testers, applied to calibrate passive leakage current testers (without isolated power supply) or active leakage current testers.
meet the requirements of jjg 843-2007 verification regulation of leakage current tester and sj/t 11383-2008 general specification of leakage current tester. the following tests are performed according to the requirements of regulations: leakage current error calibration, test voltage error calibration, input resistance measurement and input circuit time constant measurement.
applications
● calibrate active leakage current testers: use an intelligently controlled ac/dc current receiving source for calibration.
● calibrate universal leakage current testers (frequency is from 15 hz to 30 khz)
● calibrate medical leakage current testers (frequency is from 15 hz to 1 mhz)
features
● the current receiving source can track and control the changes of the power grid voltage to ensure the measurement accuracy.
● a built-in voltmeter can calibrate the error and distortion of the test voltage.
● measure the input resistance according to the requirements of regulations by the function of resistance measurement.
● measure the time constant of input circuit according to the requirements of regulations by the built-in frequency meter.
● high integration, abandons the traditional complex test methods, simple wiring and easy operation.
● good electrical protection to avoid damage caused by zeroing of the high resistance box.
electrical specifications
ac current standard source (calibrate passive leakage current testers)
output characteristics
● output range of voltage: 2 μa to 24 ma
● resolution: 0.01 % of range
● display: 5 digits
dc current standard source (calibrate passive leakage current testers)
output characteristics
● output range of voltage: 2 μa to 24 ma
● resolution: 0.01 % of range
● display: 5 digits
ac/dc current receiving source (calibrate active leakage current testers)
output characteristics
● output range of ac current: 20 μa to 24 ma
● output range of dc current: 2 μa to 24 ma
● resolution: 0.01 % of range
● display: 5 digits
other measurement
general specifications
● line power: 220 v ±22 v, 50 hz ±2 hz
● maximum power consumption: 80 va
● operating environment: +18 °c to +28 °c, 40 % to 80 % r·h, non-condensing
● storage environment: -20 °c to +70 °c, <80 % r·h, non-condensing
● dimensions (w × h × d): 405 mm × 195 mm × 370 mm
● weight: 9 kg
● warm-up: 30 minutes
● communication: rs-232 interface
selection guide