td1510 high-凯发k8国际

for more product details, please call us:0731-84930888

homeelectrical measurementdc instruments

td1510 high-precision dc calibrator
暂无内容


overview

model td1510, a wide-range, high-precision and multi-function calibrator with a built-in dc standard source. the apparatus can accurately outputs dc voltage, current, power and electrical energy with the optional accuracy classes are 0.01, 0.02 and 0.05. applied to calibrate various types of dc electric measuring instruments.

features

dc electrical parameters output: voltage is up to 1150 v, current is up to 120 a, the low-current is as low as 0.1 μa.

● value output mode: standard source and output source modes, easy to calibrate both digital meters and analogue meters.

● value adjustment method: fixed-point, rotary encoder, step and potentiometer (option).

● output switch: the output can be turned on/off arbitrarily by one-touch operation.

● statistics and analysis of data: calculate the maximum, minimum, average and standard deviation, and then analyze the stability and trends. 

● excellent human-computer interaction: large lcd touch screen display, touch and button operation modes, easy for users to measure manually.

● electric energy meter calibration: built-in a dc low-voltage standard source with an output range of ±(10 μv to 4.4 v), and standard pulse i/o interfaces to calibrate electrical energy meters by direct access or indirect access.

● transmitter measurement (option): measure the secondary dc signal of the transmitters.

● proprietary software (option): support the calibration of the checked meters by semi-automatic or full-automatic, data management and report export.

applications (class 0.05)

1

● calibrate dc voltmeters and ammeters (class 0.2 and below)

● calibrate dc power meters (class 0.5 and below)

● calibrate dc electric energy meters (class 0.5 and below) (option)

● calibrate dc transmitters (class 0.2 and below) (option)

electrical specifications

dc voltage

2

output characteristics

● output range of voltage: 20 mv to 1100 v (1150 v @ electric energy option)

● resolution: 0.001 % of range

● display: 7 digits

● ripple coefficient: <0.5 %

● protection: short-circuit protection, overload protection.

dc current

3

output characteristics

● output range of current: 0.1 μa (or 1 μa) to 120 a

● resolution: 0.001 % of range

● display: 7 digits

 ripple coefficient: <0.5 %

● protection: open-circuit protection, overload protection.

dc electric energy (option) and dc power

 output range: combination of dc voltage range and dc current (or low-voltage) range.

● accuracy: addition of dc voltage accuracy and dc current (or low-voltage) accuracy.

● input of standard electric energy pulse: 

receive high frequency electric energy pulses (≤200khz), input levels ≤ 24v

● output of standard electric energy pulse:

high frequency full scale @ 60 khz, low frequency full scale @ 6 hz.

frequency jitter: <10 μs

stability of clock frequency in 1 year: <50 ppm

● electric energy measurement error: automatically display with 0.0001 %.

5

dc small signal measurement (option)

● range: 1 v, 10 v, 2 ma and 20 ma

● measuring range: ± (0 to 12 v);± (0 to 24 ma)

● accuracy: ± (0.006 % of reading + 0.004 % of range)

● notes: applied to measure the secondary dc signal of the transmitters.

general specifications

● line power: 220 v ±22 v, 50 hz ±2 hz

● maximum power consumption: 700 va

 operating environment: 0 °c to +40 °c, 20 % to 85 % r·h, non-condensing

● storage environment: -20 °c to +70 °c, <95 % r·h, non-condensing

● dimensions (w × h × d): 550 mm × 210 mm × 465 mm

● weight: 24 kg